HP 5890 Packed Injector, refurbished
Interchangeable stainless steel liners/adapters, installed from inside the
oven, are used with the packed column inlet, and with all detectors,
depending upon the column to be installed.
Liners for the packed column inlet are available in three sizes: one for
1/8•inchcolumns, one for 1/4•inchcolumns, and one for HP Series 530.

In addition, liners for the packed column inlet are available to accept
glass inserts (discussed later) for reduced reactivity, to trap nonvolatile
residues, or for use with an HP Series 530 u capillary column.
- No liner is used with 1/4•inchpacked glass columns. The long leg of
the column fits into the inlet body, replacing the liner. Packing and
glass wool plug must be below the tip of the needle for best results.
- Metal columns are installed with a liner appropriate for the column
diameter.
- If necessary, glass columns can be installed using a metal liner
(preferably those accepting a glass insert), but this is not
recommended. There may be problems with dead volume in
connections, and preventing contact of sample with metal surfaces.
Columns and Fittings
|