Refurbished spectrophotometer Perkin Elmer Lambda 3 (190 nm – 1100 nm)

UV-Visible Spectrophotometer
- Extended wavelength range: 190 nm to 900 nm
- Scan speeds from 15 nm/min to 480 nm/min.
- Double beam optics for excellent long term stability, reference compensation and baseline correction
- Basic methods for timedrive, scan, wavelength program and concentration.
- RS-232 Serial Port for PC Software
- Stray Light (2 sec resp time): < 0.05 T at 220 nm
- Wavelenght Accuracy: ± 0.5 nm at D2 peak 656.1 nm
- Wavelenght Reproducibility: ± O.2 nm
- Bandpass: < 2 nm at D2 peak 656.1 nm* Photometric Accuracy: ±0.005 A at 1 A
- Photometric Reproducibility: ±0.002 A at 1A
- Photometric Range: -0.301 A to 3.000 A
- Photometric Stability: < 0.0005 A/h (500 nm, 1 sec resp)File Photo of similiar unit
|
|