Hood, Ductless Fume, Air Science USA, Purair 40, 96W x 27D x 47H inch

The Purair range of ductless fume hoods have been designed to provide
operator protection when using hazardous substances. A face velocity at 100 fpm
ensures containment of fumes and an alarm will alert the operator when the
airflow falls to an unacceptable level. All mechanisms in the head section of
the Purair are on the clean side of the filter, thus preventing contamination.
Switches and electrical components are totally isolated from the dirty airflow
and away from any contamination. The work area has a removable spillage tray
which can be easily cleaned. Optional integral lighting is available.
FILTER SELECTION
The main filter can be chosen from 14 different types of carbon,
which include specialty media for vapors of organics, solvents, acids,
mercury and formaldehyde. HEPA filters for particulate filtration are
also available to suit your application needs. In addition, the Purair
can be equipped with a secondary back-up filter for added protection.
This filter is required by US standard ANSI Z9.5 section 4.12 4.2.
PURAIR 10XL TYPICAL APPLICATIONS
Typical applications include weighing, chemical sampling,
histology, forensic and pharmaceutical.
EFFICIENT AND ECONOMICAL
These units are energy saving. Clean air is recirculated into the
laboratory, therefore, they do not exhaust expensive conditioned
and/or heated air into the atmosphereent of hazardous substances thus
protecting both the environment and gal requirements.
Supply 110v 60Hz single phase
(other voltages available on request)
Switches Mains on/off
Monitoring Low Airflow
All UNITS
Spillage Tray Polypropylene (optional)
Fan Centrifugal
OPTIONS
Color Back panel - Clear
Color Side panel - Clear
ELECTRICAL
- Internal lights/Safety exhaust filter/Electronic filter
- saturation alarm/Spillage tray
- Airflow (CFM) - 220
- Face Velocity (FPM) – 440
- Lighting - 2 x 15watts
- Noise (1m from unit) - <56dBA
- Width/Depth/Height (inch) - 96 x 27½ x 47¼
- Construction - Epoxy coated mild
- steel frame & head unit
- Color - White
- Color Back panel - Clear
- Color Side panel - Clear
- Spillage Tray Polypropylene (optional)
- Fan Centrifugal
- Supply 110v 60Hz single phase
- (other voltages available on request)
- Switches Mains on/off
- Monitoring Low Airflow
- Internal lights/Safety exhaust filter/Electronic filter
- saturation alarm/Spillage tray

|