Scientific Equipment Source
1895 Clements Rd., Unit 127 Pickering, ON, L1W 3V5
Canada
Tel: 905-231-0930
Fax: 905-231-0293
|
ATR-SW, automatic microprocessor controlled critical angle refractometer
This SCHMIDT+HAENSCH automatic
microprocessor controlled critical angle refractometer is designed to
measure the refractive index of liquid media independent of opacity,
viscosity and colour. It consists of separate electronic and measuring
devices to enable single and continuous measurements. Convincing design
of device from stainless steel allows horizontal and vertical
application of the sample room with a thermostatable sapphire prism.
The prism is illuminated by a 589 nm LED and the angle of total
reflection is detected by a diode array which guarantees high
resolution and stability. The ATR-SW enables to complete measurements
in a quick and easy way by using splash proofed touch-in panel. It's
new housing for the electronic offers more space for a bigger LCD
display and an alphanumeric foil keypad. Batch numbers can be entered
directly and also simple forms can be created and printed out. The
scale key offers two scales; refractive index and %Brix temperature
corrected within +10 to +50 °C range.
Different sample room doors with and without water circulation
facilities for manual or automatic sample changing complete the
versatility.
Features
- Measurement of liquid media irrespective of viscosity, opacity and colour
- Electronic and measuring devices separate
- Single and continuous measurements
- Different sample room doors
- Date and time function
- Up to 4 user--defined scales programmable
Applications
- Chemical industry :Solvents,organic polymers, destilled products,solutions of anorganic compounds.
- Food industry:Yoghurt,jam,fruit extract, syrup, honey,coffee extract,vegetable fat,albumen, chocolate, milk, baby food etc.
- Beverage:Soft drinks,fruit juice,wine,beer.
- Medical applications:Serum,blood,urine.
- Pharmaceutical /cosmetical industry:Essences, essential oils,aromas.
- Petrochemics:Oils,fats,waxes,naphtalenes,raw oils,
- varnishes,paints.
Technical Specification
- Measuring ranges RI 1.33200 -1.53200 / Brix 0 - 95%
- Resolution RI 0.00002 / Brix 0.02 %
- Precision RI ± 0.00004 / Brix ± 0.04 %
- Working temperature +5 °C up to +50 °C
- Display Illuminated 16 character LCD
- Prism Sapphire,thermostatable
- Light source 589nm LED
- Data output RS 232 C and parallel interface, USB optional
- Dimensions Electronic device 185x360x165mm Measuring device 160x195x155mm
- Weight 8 kg (complete unit)
|
|